Fluke Calibration
Fluke 2271A Industrial Pressure Calibrator
Model No :
Fluke 2271A
Part Code :
Please contact us for final configuration

Overview

The Fluke 2271A features pressure ranges from -100 kPa to 20 MPa (-15 to 3000 psi), which covers the requirements of a wide range of gauges and sensors. A built-in electrical measurement module (EMM) with HART capabilities enables you to perform closed loop, fully automated calibration on 4-20 mA devices such as smart transmitters, gauges and switches.
The EMM supplies 24 V dc loop power for measuring mA and V dc. It has a built-in 250 Ohm resistor that can be toggled on or off, eliminating the need to have an external resistor to enable HART communications.

Features
- Calibrate a wide range of gauges and sensors with a single instrument
- Two levels of accuracy, 0.01% reading or 0.02% FS
- Wide measurement range from -100 kPa to 20 MPa (-15 to 3000 psi)
- Removable pressure measurement modules make it easy to change or add measurement ranges
- Integrated electrical measurement module with HART communication provides a complete solution for calibrating pressure transmitters
- Built-in dual test ports enable you to connect multiple devices under test (DUTs)
- Localized graphical user interface in choice of ten languages

Major Applications
- For calibrating a wide variety of pressure gauges and sensors
Technical Specifications
Pressure Limits Working Pressure Range : -97.90 kPa (-14.2 psi) gauge to 20 MPa (3000 psi) gauge
- Supply Port : 23 MPa (3300 psi) gauge
- Test Port : 20 MPa (3000 psi) absolute
- Reference Port : 115 kPa (17 psi) absolute
- Vent Port : 150 kPa (22 psi) absolute
- Supply Gas Type :
• Clean Dry Air or Nitrogen (Industrial Grade, 99.5 %)
• Maximum Particulate Contamination : =1.25 micrometer (50 microinches)
• Maximum Moisture Content : -50 °C dew point
• Maximum Hydrocarbon Content : 30 ppm
- Vacuum Supply :
• ->50 liters per minute capacity with Auto Vent feature
• Exhaust gas will pass through the vacuum supply. Appropriate protections should be taken when operating at higher pressures.
- Source :
• Loop Compliance Voltage : ³24 V dc @ 20 mA (without 250 W built in resistor)
³19 V dc @ 20 mA (with 250 W built in resistor)
- Switch Testing :
• Current Limits : £4 mA
• Conditions : Closed >1.5 mA, Open <0.5 mA

General Specifications
- Power requirements : 100 V ac to 240 V ac, 47 Hz to 63 Hz
- Fuse : T2A 250 V ac
- Max power consumption : 100 W
Environment
- Operating Ambient Temperature Range : 15 °C to 35 °C
- Storage temperature : -20 °C to 70 °C
- Relative Humidity : Operating: <80 % to 30 °C, <70 % to 35 °C Storage: <95 %, non-condensing
- HART Communication and Functions
- HART Modes : HART mA measurement with 24 V (Loop)
- Automatic HART Detection : HART connected with automatic polling
- HART Selectable Resistor : Built-in 250 W loop resistor, Selectable ON/OFF
- HART Commands : HART universal and common practice commands (no device specific commands)
- Write Protection : HART Write enable/disable